TOSS
   Übersicht über weitere verfügbare Pyranometer

     Technische Daten

Specifications CMP 11 CMP 21 CMP 22
ISO 9060:1990 Classification Secondary Standard Secondary Standard Secondary Standard
Response time (95 %) < 5 s < 5 s < 5 s
Zero offsets
(a) thermal radiation (200 W/m²)
(b) temperature change (5 K/hr)
< 7 W/m²
< 2 W/m²
< 7 W/m²
< 2 W/m²
< 3 W/m²
< 1 W/m²
Non-stability (change/year) < 0.5 % < 0.5 % < 0.5 %
Non-linearity (0 to 1000 W/m²) < 0.2 % < 0.2 % < 0.2 %
Directional error (up to 80 ° with 1000 W/m² beam) < 10 W/m² < 10 W/m² < 5 W/m²
Temperature dependence of sensitivity < 1 %
(-10 °C to +40 °C)
< 1 %
(-20 °C to +50 °C)
< 0.5 %
(-20 °C to +50 °C)
Tilt error (at 1000 W/m²) < 0.2 % < 0.2 % < 0.2 %
       
Other specifications      
Sensitivity 7 to 14 μV/W/m² 7 to 14 μV/W/m² 7 to 14 μV/W/m²
Impedance 10 to 100 Ω 10 to 100 Ω 10 to 100 Ω
Level accuracy 0.1 ° 0.1 ° 0.1 °
Operating temperature -40 °C to +80 °C -40 °C to +80 °C -40 °C to +80 °C
Spectral range (50 % points) 285 to 2800 nm 285 to 2800 nm 200 to 3600 nm
Typical signal output for
atmospheric applications
0 to 15 mV 0 to 15 mV 0 to 15 mV
Maximum irradiance 4000 W/m² 4000 W/m² 4000 W/m²
Expected daily uncertainty < 2 % < 2 % < 2 %
Recommended applications Meteorological networks, PV panel and thermal collector testing, materials testing Meteorological networks, reference measurements in extreme climates, polar or arid Scientific research requiring the highest level of measurement accuracy and reliability

Impressum | Datenschutzerklärung | Copyright © TOSS GmbH Potsdam
Senden Sie E-Mail mit Fragen oder Kommentaren zu dieser Website an: TOSS GmbH
Stand: 02.07.2022